Annual calibration and quality assurance for intraoral, panoramic, and CBCT units. We verify kVp and mA output, check beam alignment, measure half-value layer, and test the exposure timer. Certificate is signed off for Ministry of Health inspections.

Calibration and safety testing for intraoral, panoramic, and CBCT systems against IEC 60601-2-65 and the manufacturer service manual.
kVp measured at the rated setting (typically 60-70 kVp for intraoral), mA verified, exposure time checked against the timer setting. Drift outside ±10% gets corrected before sign-off.
Half-value layer measured, beam alignment to receptor, leakage from the tube head at 1 m, warning light and door interlock tested. Findings logged on a dosimeter report.
Signed certificate for the radiological safety body, with raw measurements attached. Accepted by the Ministry of Health in Albania and the AKR in Kosovo for the annual audit.
Calibration paired with cone, collimator, and timer service. Most clinics combine the visit with a generator capacitor check and door seal inspection on the X-ray room.
How this same expertise plays out on a real dental clinic project.

Recent articles on dental practice design, equipment and patient flow.

How cutting-edge design trends are transforming dental practices, improving patient outcomes, and boosting clinic efficiency through 2026.

How thoughtful design choices can significantly improve patient satisfaction and reduce dental anxiety.
How a calibration visit runs from booking to certificate.
Most clinics book 2-3 working days ahead. Visits run outside patient hours so the unit can be taken offline without losing chair time.
Tube head, cone, arm, sensor or panoramic receptor, exposure switch and warning lights checked. Any wear or damage is noted before electrical tests start.
Calibrated dosimeter and kVp meter set up at the focal spot distance. kVp, mA, exposure time and dose measured at each preset, then trimmed to the manufacturer spec.
Half-value layer measured, leakage from the tube head checked at 1 m, beam alignment to receptor verified, and door interlock and warning light tested.
Calibration certificate produced with all readings attached. Copy left at the clinic, copy filed with the radiological safety authority where required.